For best results dilute the stock single-element antimony standard solution 1.000 gram per litter (CPA No C050100) to levels appropriate to the operating range of the instrument using. The working solution must be freshly prepared.

A calibration blank is used to establish the analytical calibration curve; the laboratory blank is used to assess possible contamination from the sample preparation procedure and to assess spectral background.

Calibration blank—Consists of 1% (v/v) nitric acid in high-purity analytical water.

 

q       Recomended Parameters (Varian instruments)

 

Lamp Current                  10 mA

Spectral Bandwidth         0.2 nm

Wavelenght                     217.6 nm

Maximum Ash Temp.      800 oC (Nitric acid medium)

 

For analyse of antimony by Graphite Furnace using of Palladium-Magnesium modifier (CPA No MPD01) is recomended. Maximum Ash Temperature 1000oC with chemical modifier. Typical response (Argon): 20 μl of 50 ppb Sb solution + 5 μl modifier gives about 0.45 Abs.

 