For best results dilute the stock single-element antimony standard
solution 1.000 gram per litter (CPA No C050100) to
levels appropriate to the operating range of the instrument using. The working
solution must be freshly prepared.
A calibration blank is used to establish the analytical calibration
curve; the laboratory blank is used to assess possible contamination from the
sample preparation procedure and to assess spectral background.
Calibration blank—Consists of 1% (v/v) nitric acid in high-purity
analytical water.
q
Recomended Parameters
(Varian instruments)
Lamp Current 10
mA
Spectral Bandwidth 0.2 nm
Wavelenght 217.6
nm
Maximum Ash Temp. 800 oC (Nitric acid medium)
For analyse of antimony by Graphite Furnace
using of Palladium-Magnesium modifier (CPA No MPD01) is recomended. Maximum
Ash Temperature 1000oC with chemical modifier. Typical response
(Argon): 20 μl of 50 ppb Sb solution + 5 μl modifier gives about 0.45
Abs.