For best results dilute the stock single-element cadmium standard
solution 1.000 gram per litter (CPA No C010100) to
levels appropriate to the operating range of the instrument using. The working
solution must be freshly prepared.
A calibration blank is used to establish the analytical calibration
curve; the laboratory blank is used to assess possible contamination from the
sample preparation procedure and to assess spectral background.
Calibration blank—Consists of 1% (v/v) nitric acid in high-purity
analytical water.
q
Recomended Parameters
(Varian instruments)
Lamp Current 4
mA
Spectral Bandwidth 0.5 nm
Wavelenght 228.8
nm
Maximum Ash Temp. 300 oC (Nitric acid medium)
For analyse of cadmium by Graphite Furnace using
of Ammonium Phosphate modifier (CPA No MNH03) is recomended. Maximum Ash
Temperature 600oC with chemical modifier.
Typical response (Argon): 10 μl of 2 ppb
As solution + 5 μl modifier gives about 0.2 Abs.