For best results dilute the stock single-element cadmium standard solution 1.000 gram per litter (CPA No C010100) to levels appropriate to the operating range of the instrument using. The working solution must be freshly prepared.

A calibration blank is used to establish the analytical calibration curve; the laboratory blank is used to assess possible contamination from the sample preparation procedure and to assess spectral background.

Calibration blank—Consists of 1% (v/v) nitric acid in high-purity analytical water.

 

q       Recomended Parameters (Varian instruments)

 

Lamp Current                  4 mA

Spectral Bandwidth         0.5 nm

Wavelenght                     228.8 nm

Maximum Ash Temp.      300 oC (Nitric acid medium)

 

For analyse of cadmium by Graphite Furnace using of Ammonium Phosphate modifier (CPA No MNH03) is recomended. Maximum Ash Temperature 600oC with chemical modifier.

Typical response (Argon): 10 μl of 2 ppb As solution + 5 μl modifier gives about 0.2 Abs.

 