For best results dilute the stock single-element boron standard solution
1.000 gram per litter (CPA No C005100) to
levels appropriate to the operating range of the instrument using. The working
solution must be freshly prepared.
A calibration blank is used to establish the analytical calibration
curve; the laboratory blank is used to assess possible contamination from the
sample preparation procedure and to assess spectral background.
Calibration blank—Consists of 1% (v/v) nitric acid in high-purity analytical
water.
q
Recomended Parameters
(Varian instruments)
Lamp Current 20
mA
Spectral Bandwidth 0.2 nm
Wavelenght 249.8
nm
Maximum Ash Temp. 1000 oC (Nitric acid medium)
Typical response (Argon): 10 μl of 3.5 ppm
B (+30 μl Ba 1 g/l CPA NoC006100) solution gives about 0.1 Abs.