For best results dilute the stock single-element arsenic standard
solution 1.000 gram per litter (CPA No C003100) to
levels appropriate to the operating range of the instrument using. The working
solution must be freshly prepared.
A calibration blank is used to establish the analytical calibration
curve; the laboratory blank is used to assess possible contamination from the
sample preparation procedure and to assess spectral background.
Calibration blank—Consists of 1% (v/v) nitric acid in high-purity
analytical water.
q
Recomended Parameters
(Varian instruments)
Lamp Current 5
mA
Spectral Bandwidth 1.0 nm
Wavelenght 193.7
nm
Maximum Ash Temp. 300 oC (Nitric acid medium)
For analyse of arsenic by Graphite Furnace
using of Palladium-Magnesium modifier (CPA No MPD01) is recomended.
Maximum Ash Temperature 800oC with chemical modifier.
Typical response (Argon): 10 μl of 100 ppb
As solution + 5 μl modifier gives about 0.3 Abs.