For best results dilute the stock single-element aluminium standard
solution 1.000 gram per litter (CPA No C002100) to
levels appropriate to the operating range of the instrument using. The working
solution must be freshly prepared.
A calibration blank is used to establish the analytical calibration
curve; the laboratory blank is used to assess possible contamination from the
sample preparation procedure and to assess spectral background.
Calibration blank—Consists of 1% (v/v) nitric acid in high-purity
analytical water.
q
Recomended Parameters
(Varian instruments)
Lamp Current 10
mA
Spectral Bandwidth 0.5 nm
Wavelenght 309.3
nm
Maximum Ash Temp. 1400 oC (Nitric acid medium)
Typical response (Argon): 20 μl of 25ppb Al
solution gives about 0.3 Abs.
For complex samples and/or matrixes
Phosphates-Magnesium modifier (CPA No MNM05) is recommended. Use this solution for addition to blanks, calibration standards and
samples.